VOLTAGE INPUT |
CURRENT INPUT |
EXTERNAL
SENSOR INPUT |
POWER |
| Ranges |
30 V, 150 V, 300 V, 600 V |
1 A, 5 A, 10 A, 20 A |
50 mV, 250 mV, 500 mV, 1.0
V |
--- |
| Crest Factor |
2.7 @ full scale input |
1.7 @ full scale input |
2.4 @ full scale input |
--- |
| Impedance |
2 megohm |
12 milliohm |
17 kilohm |
--- |
| Display Range |
5 digits with 10 mV resolution |
5 digits with 1 mA resolution |
5 digits with 0.1 mV resolution |
6 digits with 1 mW resolution |
| ACCURACY |
| DC |
±(0.1%
Reading + 0.2% Range) |
0.4% of VA range |
| 5 Hz - 500 Hz |
±(0.1%
Reading + 0.1% Range) |
0.2% of VA range |
| 500 Hz - 10 kHz |
±((0.015
× F(kHz)% Reading) + 0.3% Range) |
0.6% of VA range |
| 10 kHz - 100 kHz |
0.6% + (0.03% × F(kHz)% of VA range) |
| >100 kHz |
N/A measurement band limited
to DC-100 kHz |
| FREQUENCY |
| Source |
V1,A1, V2,A2, V3,A3,
LINE or EXT
(For Vx and Ax, source input must be >10% of range
setting) |
| Accuracy |
20 to 500 Hz 0.05% |
| EXT. Input |
BNC non-isolated, earth ground
referenced |
| Input Impedance |
100 kilohm |
| Voltage Level |
TTL / CMOS |
| Maximum Voltage |
50 V |
|